Is it actually the peptide on the label?
A peptide has an exact mass. We ionize the sample, measure the mass-to-charge of what comes off the column, work back to the molecular mass, and compare it with the expected mass for the named peptide. Then we fragment it to confirm the sequence, not just the weight.
How it works
- 01Dissolve and separateA few micrograms are dissolved and run through a reversed-phase column so the peptide reaches the detector apart from salts and impurities.
- 02IonizeElectrospray puts charge on the molecule. Peptides pick up several protons, so one peptide shows up at several m/z values: [M+2H]²⁺, [M+3H]³⁺ and so on.
- 03Measure and deconvoluteThe instrument records each m/z to a fraction of a Dalton. We read the charge state from the spacing of the isotope peaks and compute the neutral mass. For BPC-157, [M+2H]²⁺ at 710.8 gives 1419.5 Da.
- 04Confirm the sequenceThe ion is fragmented and the fragment ladder is matched to the expected sequence. Two peptides can share a mass; they rarely share a fragment pattern.
What the result means — and what it doesn't
Match means the measured mass agrees with the expected mass within tolerance and the fragment pattern fits the sequence. Mismatch means it does not — the result names the closest known peptide when there is one (a semaglutide vial that measures as liraglutide, for instance). Identity says nothing about how much is there or how pure it is; those are separate tests on the same page.
Why another lab might report a different number
- Monoisotopic versus average mass: the two differ by about 1 Da for a 1,400 Da peptide. We report which we used.
- Charge-state misassignment turns a correct [M+2H]²⁺ into a wrong mass. We print the observed m/z and charge so the arithmetic is checkable.
- Salt adducts (TFA, acetate, sodium) shift peaks. Our tolerance window and adduct handling are stated on the result.
- Mass alone cannot separate isomers or some analogs; a lab that skips fragmentation may call a match where we would not.
That's why every result here prints the method and the limit next to the number. The full science article on why tests vary.
Methods described are Vialidate's own and are published so results can be compared with other laboratories' on equal terms. Nothing on this page is medical advice. ISO 17025 accreditation in progress.