LC-MS/MS · targeted MRM · +$50

Is there fentanyl, or a fentanyl analog, in the vial?

A targeted mass-spectrometry method that looks for fentanyl and a list of analogs by their specific precursor-to-fragment transitions, at half a nanogram per millilitre. It is cheap because the identity run already has the sample on the instrument.

LC-MS/MS MRM · VL-2026-[00404a]NOT DETECTED
spiked control, 0.5 ng/mLm/z 337.2 → 188.1sample: flat
Not detected < 0.5 ng/mL
Illustrative readout — real lot spectra replace this when the registry opens.

How it works

  1. 01
    Same sample, second method
    An aliquot of the identity-run solution is injected on a short, targeted method.
  2. 02
    Watch specific transitions
    The instrument monitors fentanyl's precursor ion (m/z 337.2) fragmenting to m/z 188.1 and 105.1, plus the transitions for each analog on the list.
  3. 03
    Compare with a spiked control
    A control at the detection limit runs with every batch. If it doesn't show, the batch is invalid.
  4. 04
    Report
    Not detected below 0.5 ng/mL in the reconstituted sample, or the value and the analog identified.

What the result means — and what it doesn't

Not detected means none of the listed compounds produced a signal above 0.5 ng/mL in the tested solution. The analog list is published and grows. A detection is reported with the compound name and value, and the vendor is notified before publication like any other result.

We will not say: "drug free" or "screened for all opioids". We screen for a named list, at a stated limit.

Why another lab might report a different number

  • Analog coverage: a method that watches ten analogs will miss the eleventh. The list is on the result.
  • Reconstitution volume changes the per-mL number, as with endotoxin.
  • Immunoassay test strips have different cross-reactivity and a higher limit; they are a different tool.

That's why every result here prints the method and the limit next to the number. The full science article on why tests vary.

Methods described are Vialidate's own and are published so results can be compared with other laboratories' on equal terms. Nothing on this page is medical advice. ISO 17025 accreditation in progress.